Series workbench#

The Series tab of the Analysis workbench follows peaks across several spectra without a script: a stacked plot of the spectra, a trend for each region you track, and a table of the points. Use it for a reaction followed over time, a titration, a temperature or concentration series, or an arrayed T1, T2 or DOSY experiment.

Opening scans as a series#

There are three ways in.

  • Open as series in the right-click menu of a group in the Data Browser. The group's 1D spectra become the working set, oldest first, and the menu item shows how many spectra it will use. It needs at least two 1D spectra, or one arrayed acquisition (T1, T2, DOSY or kinetics) in the group; otherwise it is disabled.
  • Open as series on a single arrayed scan, which uses that scan's increments as the series.
  • Open in Series on the T1 or T2 page's fit table, which opens the scan the fit came from with one region per fitted window. The button asks which model: Exponential (the page's own), Stretched exponential or Bi-exponential, for decays with a spread of rates or two components. It is disabled until that scan is loaded and the fit has windows.

If Analysis already has other scans open, you are asked to confirm Replace and open. If you reopen the same members, your regions and settings are kept; with different members they start fresh, but your model choice stays. Scans in the group that are not 1D spectra or arrayed acquisitions are left out, and the toolbar counts them (hover for why).

The toolbar#

Control What it does
Against The x axis: Time (acquisition time, with a Unit of Auto, s, min or h), Entered value (you name it and give a unit, for example pH) or Arrayed axis (offered for arrayed scans).
Align Not aligned, To each sample's reference (internal standard if present, else solvent residual), or on the centroid of one region. Spectra that cannot be aligned are counted and listed on hover.
Fit Best fit (AICc), No fit, Zero order, First order, Second order, Exponential, Stretched exponential (a decay with a spread of rates: reports the stretching exponent β and the mean time), Bi-exponential (two components: reports a fast and a slow time constant) or Sigmoid (titration).
Quantity What the trends plot: Integral, Mole fraction, Concentration, Conversion or Yield. Everything except Integral needs species.
Follow this group Keeps the series current while scans arrive; see Following a group.
Export Export trends and fits as CSV.
Add to report Appends a series section (stacked spectra, trend, fit table) to the report template.

Stacked plot#

The spectra are drawn stacked, first at the bottom. Offset sets the vertical step as a fraction of the tallest peak. Click a peak to track it, or right-click and drag to draw a region. Regions you draw appear in the Trends list.

The side panel has four tabs: Trends, Species, Scheme and Spectra.

Trends lists one row per region with its fit result and the R². Suggest regions tracks the tallest distinct peaks, each labelled by its apex; Clear all removes every region and can be undone. Open a region to edit its Label, High (ppm), Low (ppm) and Measure (Integral, Height or Position (ppm)), or to remove it. A region can also be marked as the internal standard, so the other integrals are divided by it, or as the alignment reference. The row shows the fit parameters with their uncertainties (rate constant, half-life, plateau, midpoint, and so on, depending on the model); with Best fit (AICc) it says which model won and which were not considered, and why.

When a region came from a T1 or T2 page, the row says whether it uses the same window as that page and offers Use the page's window if it does not.

Spectra is the points table. Each row has a Use check box: untick it to leave that spectrum out of all trends and fits. With Entered value you type the x value for each spectrum; spectra without one are left out and a message counts them. When all spectra share one time or value, the panel says there is nothing to plot against and offers Use spectrum order or Enter values.

Species and quantities#

The Species tab names the participants of a reaction and ties each to the regions that measure it.

  • A species has a Name, a Role (reactant, product, intermediate, by-product or internal standard) and, for a product, the Moles per mole of the first reactant at full conversion.
  • Each region assigned to a species carries its Nuclei count, so integrals of signals with different proton counts compare as amounts. Without it a region counts as one nucleus, and the panel says so.
  • A species is linked to a Sample component, or defined in the series only. Add as component records a series-only species in the sample and links it.

The Quantity control then plots, for each species:

Quantity Definition
Mole fraction The species' amount over the amount of all species except the standard, in percent.
Concentration The amount against the internal standard, in mol/L. The standard's concentration is read from its sample component, the one with the internal-standard role; it is not entered again here. It uses the internal-standard equation of qNMR.
Conversion 1 − n(reactant, t) / n(reactant, first point), in percent, for each reactant.
Yield n(product, t) / (n(first reactant, first point) × moles per mole), in percent, for each product.

All of these run through the same fits as the integrals, with the same uncertainties. The first point of the series is the start of the reaction.

Beside every species quantity the panel shows the verdict of the qNMR acquisition checks for the first spectrum (relaxation delay, signal-to-noise, digital resolution and the others). Benchtop kinetics often use a short delay, which biases the numbers; when a check fails, the panel states that the quantity is not quantitative here and names the check. The uncertainty of the standard's own concentration is not included.

Reaction scheme#

The Scheme tab draws the reaction: the species with the Reactant role, an arrow, and the species with the Product role, each structure drawn by the same renderer as the spectrum overlay. A species is drawn from its SMILES, or from the sample component it is linked to when it has none. Intermediates, by-products and the internal standard are not drawn. Under each structure are its name and the latest value of the chosen quantity (for example conversion 85 %).

Conditions is free text written over the arrow, one row per line: reagents, solvent, temperature. It is saved with the series. A species that cannot be drawn is left out of the scheme.

Following a group#

Follow this group appends every scan saved into the group to the working set, whether it comes from an acquisition or an import, and the trends refresh as each one arrives. The setting is saved in the project and applied by the backend, so the series is up to date when you open the project later, with no window open during the run. It is available when all series scans are in one group.

Reaction report#

Add to report also appends the species table, the reaction scheme, the plot of the chosen quantity, the kinetics table and the basis of the figures when species are defined. The built-in Reaction Report template uses the same fields:

Field Content
{{series.species}} Species, roles, regions and whether each is linked to a sample component.
{{series.scheme}} The reaction scheme as a picture, with the final conversion of each reactant and yield of each product under its structure. Empty when no species has a structure.
{{series.quantity}} Plot of the chosen quantity for every species.
{{series.conversion}}, {{series.yield}} Conversion of the reactants and yield of the products, whatever the workbench shows.
{{series.kinetics}} Fit parameters (rate constants, half-lives) per species.
{{series.caveats}} The notes and the acquisition verdict the quantity rests on.
{{series.excluded}} The points left out by choice.

An approved report keeps the values it was approved with.

While loading#

Reading the spectra can take a while. A floating list at the top right of the plot shows each member as queued, loading, loaded or failed (hover a failure for the reason), with a count such as "Reading spectra 3 / 12". The trends appear once the spectra have been read.

Note

Without Entered value, spectra with no acquisition time are left out of the trends; a message gives the count.