Spectrum view#
The spectrum view is shown on Acquisition, the Data Browser preview, the relaxation pages, custom pages and plugin pages. It draws the data, and its right-hand tool sidebar is the way into peak analysis, assignment, qNMR, compound matching, elucidation, DOSY, phase correction and solvent handling.

Choosing what is drawn#
A scan can be shown in the time domain or the frequency domain, and a 2D scan can be drawn several ways. The view chooses a representation automatically from the data; choose a different one from the representation menu.
| Representation | Data |
|---|---|
| 1D Signal | Time-domain FID |
| 1D Spectrum | Frequency-domain spectrum |
| 2D Signal Heatmap | Time-domain 2D signal |
| 2D Spectrum Heatmap | Frequency-domain 2D spectrum |
| 2D Contour with Projections | Contour plot with 1D projections on both axes |
| 3D Surface | Orbitable surface, for signal or spectrum |
| Ridge Plot | Stacked 1D slices |
On a 2D plot, the 2D Controls panel shows or hides the axis projections and reports the cursor position as F1, F2 and intensity.
Tool sidebar#
The buttons down the right-hand edge are the analysis tools. They are grouped — peaks and assignment, identification, then general tools — with a divider between groups.
Buttons appear and disappear depending on what is on screen. A tool you used earlier is not missing; it does not apply to the scan you are looking at now. The rules are:
| Tool | Appears when |
|---|---|
| Peaks | A spectrum is shown, and it is not a ridge plot. 1D and 2D both. |
| Peak assignment | A 1D spectrum. A crosspeak names an atom pair, which the 1D panel has no room for. |
| Cross-peak assignment | A 2D spectrum, except a J-resolved plane. On a J-resolved plane, F1 carries coupling rather than a second shift, so its cross-peaks name no atom pair. |
| qNMR | A 1D spectrum. |
| DOSY | A pseudo-2D scan whose experiment type is DOSY — the indirect axis has to carry gradient strength for there to be anything to invert. |
| Identify compounds | Any spectrum that is not a ridge plot. |
| Structure elucidation | Any spectrum that is not a ridge plot, if your licence includes elucidation. |
| Processing history | Whenever the plot has data. |
| Scan review | The Acquisition plot, spectrum and signal. An operator sees it only if the operator policy opens the Acquisition page. |
| Phase correction | A 1D spectrum. |
| Show molecule | Any spectrum — the overlay is a drawing on top of the plot, so it applies to a 2D plane as much as to a trace. The structure is drawn as you drew or imported it, with its own layout. Hover the overlay for Show atom numbers (labels follow the app-wide Atom numbering setting, conventional ring locants or sequential structure order; each click steps through off, assigned and hydrogen-bearing atoms plus ring positions, and every atom) and Edit structure; reports show the same drawing. |
| Reference to solvent | A 1D spectrum whose sample records a solvent and a reference frequency. A notification reports the shift applied and offers Undo (or Ctrl/Cmd+Z). |
| Subtract the solvent line | A 1D spectrum. Shown but disabled when the sample records no solvent. |
| Ridge whitewash, Ridge flip | A ridge plot. |
Use Show / Hide Buttons at the foot of the sidebar to hide tools you never use.
Several tools open a panel. A panel can be docked into the window or popped out into a separate window, which is useful for keeping a peak table beside the spectrum on a second monitor.

Peaks#
Peak detection offers two modes, Learned (ML) and Classical. Learned (ML) needs a phased spectrum. With Auto threshold on, the noise level is measured from the spectrum and the manual threshold and prominence controls are not used. Switch from Detect to Pick by hand to place peaks by hand — click the spectrum to add one, Alt+click to remove.
Conservative multiplets is on by default. It reports ambiguous overlapping signals as generic m groups rather than assigning weak patterns. Turn it off to include uncertain patterns. The checkbox is beside Infer missing lines below the detector controls; changing it reanalyzes the spectrum and the choice is saved with the project.
The table reports shift, height, area, multiplet pattern, J couplings and FWHM, and can be downloaded as CSV. Overlays are toggled separately: peak markers, integrals, shift clues and the J-coupling tree.
Set a peak as the reference integral (1.00) from its row; the remaining integrals are then reported against it.
Note
Clearing peaks asks for confirmation and removes every peak, hand-placed ones included, as one step that Undo (or Ctrl/Cmd+Z) brings back. Auto-detected peaks can also be recovered with Re-analyze; hand-placed ones can only be restored by undoing.
Noise and signal-to-noise#
The peak table has an S/N column, one value per line. Hovering the column header names the convention and where the noise was measured. The noise comes from the quietest stretch of the trace, found automatically and kept clear of solvent-suppression bands, unless you set a region yourself.
Set a region with the Noise and S/N tool (key N, 1D only): right-drag across a stretch that holds only noise, then press Enter or Use as noise. Esc drops a span that is waiting, then leaves the tool. The region is saved with the scan, stays put when you change the chemical-shift reference, and is one step for Undo. The chip under the plot shows σ and the region; its delete button returns to the automatic estimate. The same figure is the signal-to-noise the qNMR checks read.
The convention is a project choice, under the detector controls in the Peaks panel (Signal-to-noise as):
| Convention | Formula |
|---|---|
| Height / 2σ (default) | Line height over twice the noise rms, the Bruker sino figure. For Gaussian noise it equals 2.5 × height over the peak-to-peak noise. The rms is taken after removing a straight trend from the noise region. |
| Height / σ | Line height over the noise rms. |
| 2 × height / peak-to-peak | The pharmacopoeial form. |
Integral regions: sum, split, merge, baseline#
Ctrl/Cmd-click or Shift-click rows of the integration table to select several regions. The Σ row under the table adds the integrals of the selected regions as the table reads them (every region when fewer than two are selected). Merge (key J) joins the selected regions into one that spans them all; Split (key S) cuts the selected region at the ppm in the field, the midpoint by default. Both are one undo step.
Slope and Bias subtract a straight baseline from a 1D region before it is integrated: the line is bias + slope × (ppm − region midpoint), in trace intensity units, so a plain bias lowers the whole region. Level to edges fits the line through the trace at both edges of the region. The table, the plot labels and qNMR all integrate with the line removed.
Multiplets: J values and define by range#
The signal menu (right-click a signal on the plot or in the peak table) has fields for the multiplicity and for J in Hz. Apply writes both as one undo step. A pattern whose letters disagree with the number of J values (a dd with one J) shows a warning and still applies. Integral = N in the same menu sets the scan's integral scale; the other integrals stay in proportion to it.
The Define multiplet by range tool (key M, 1D only) groups the lines inside a range you right-drag into one multiplet with pattern m. If the detector found no line there but the range holds a clear maximum, a line is added at it first; a range with neither is refused with a message.
Phase correction#
The phase panel offers zero-, first- and second-order correction on 1D spectra, and zero- and first-order correction on 2D planes. It docks beside the plot by default. Each change you settle on (releasing a slider, entering a value, Auto with the ACME method, or Reset) is applied at once as one undo step, so Undo steps back through them. If a correction cannot be applied the panel stays open and shows the reason.
By default the first-order term pivots about the edge of the spectrum, so changing it moves every line. Set pivot and then click the spectrum to pivot about that point instead: a dashed line marks it, ph0 becomes the phase at the pivot, and the line there keeps its phase while ph1 is adjusted. Moving the pivot never changes the phase on screen. The exponential method has no pivot.
A correction is previewed before it is applied. While a preview is live the peak list is cleared, because it described the previous spectrum; apply or discard the change and detection re-runs.
Which way up: DEPT, APT and inversion recovery#
In these experiments some lines point down by design, so automatic phasing keeps both signs and has to decide which way up the whole spectrum goes. It uses whatever the sample provides: a declared internal reference or a known deuterated solvent line, the sample's structure (CH₂ points down and CH/CH₃ up in a DEPT-135; the reverse for CH/CH₃, with C/CH₂ up, in an APT), or, failing both, the tallest line, shown up.
When the orientation rests on a strong reference line, it is applied silently. Otherwise the page raises a warning, because the choice is less certain.
The warning names the basis that was used and offers Flip. Flip is also next to Auto in the phase panel; there it is available when no correction is pending.
Flip turns the spectrum over with a 180° zero-order step that is stored with the scan's own processing. It survives reprocessing, shows in the processing history and can be undone. Once a scan has been flipped, its orientation counts as your decision and the warning no longer appears.
Phase of several traces#
When the open scan is on a stacked page (an overlay or stack of several scans), the phase and Processing panels start with a trace selector: a list of the page legend's traces, with their legend names and colours. Choosing one makes it the main scan, as the legend's crown does, so the panel then acts on that trace and its phase is stored in that scan's own processing, independent of the others. Alt+] and Alt+[ step to the next and previous trace of the page (within the open project).
Two commands in a page's context menu in the Pages panel act on every other trace of the page at once. Apply this phase to other traces sets the open trace's manual phase on each of them; Add this phase to other traces' own phase adds its values, as an offset, to the phase each trace is processed with (this is the only form offered for an automatic phase, whose stored values are offsets). Auto-phase every trace gives each trace automatic phasing on its own data. Each command is one undo entry, which puts every changed trace back to the processing it had. Planes, and processed imports for automatic phasing, are skipped with a reason.
Processing history#
Every operation applied to a scan is recorded — what ran, with which parameters, from where (pre-processing, the Processor, the library, generation, or numeric operations), by whom, and the data shape before and after. Open it from the sidebar to see how the spectrum on screen was arrived at.
The same panel switches between Acquisition, Processing and Sequence. Acquisition lists what the scan recorded: the instrument and conditions, dimensions, the processing the vendor stored (which is not what this app applied; that is Processing), the vendor's phase and referencing per dimension, shims, parameter lists, audit trails and every vendor parameter, searchable. All of it is read-only. Sequence draws the pulse sequence stored with a scan acquired on an instrument, as a timeline of channels; a scan imported from vendor files shows the pulse program source the file carries. A scan without an acquisition record has no Acquisition sections.
Scan context#
A line above the plot, and under the scan in the Data Browser, names the scan's sample, the number of generated reports that included it, its review state and its assignments. Each part opens what it names; hovering the scan or the sample shows a card.
Scan review#
Scan review is the review rail for the scan the plot shows: where its review stands, the items in review, the discussion, the history and the decision bar, as described under Signing and review. A comment can anchor on the peak selected on the plot. The panel follows the plot: on a page switch it moves to the new scan straight away, before that scan has finished loading. On a page that combines several scans, they are listed in the rail's checklist to switch between. Opening a scan from the Reviews page opens this panel. The sidebar button carries a badge with the scan's open comment count, coloured by its review state.
Copying and saving the plot#
With the plot focused, Ctrl+C (⌘C on macOS) copies the plot to the clipboard as a picture (PNG) together with a vector drawing (SVG text). Programs that take a picture paste the PNG; vector editors and SVG-aware documents take the SVG. The download button saves a real SVG: each trace is a single path with every data point in it, nothing thinned, and axes, ticks, labels, the legend, annotations and overlays are drawn as lines and text. A 2D plane (heatmap) is the exception: it stays an embedded picture inside the SVG.
Whether a document accepts the pasted SVG depends on the program; Word, PowerPoint and Keynote are better served by the saved SVG file than by paste.
Ctrl+P (⌘P) prints the current view: a one-page report with the spectrum as shown, opened in the report preview. See Print the current view.
Pages#
Each scan opens as a page; scans can be combined on one page as an overlay, a stack, side by side or above and below. The page keeps how it is drawn: layout, scaling, spacing, line styles and the window of the plot. A zoom or pan of the main plot is saved on its page after it settles, and the page opens at that window again with the project or the .nuclei file; resetting the zoom clears it. A window saved in one frequency unit is converted when the plot shows another.
Free page#
The Free page layout gives every pane its own box on the page, in fractions of the page, so panes can be placed anywhere, sized independently and overlap. A page holds any number of panes, and spectra and planes can share it.
- Add plot in the lower left of the canvas asks which scans to add; a scan dragged from the Data Browser onto the canvas adds a pane at the drop point, and onto an existing pane overlays it there. Scans stored with the same box share one pane.
- A pane moves by its handle on the top edge and resizes by its edges and corners, all with the left mouse button. The right mouse button on the handle opens Bring to front, Send to back and Remove pane; removing a pane does not delete the scan. Tile panes lays the panes out in a grid again. Page panes are always above scan panes.
- A page dragged from the Pages panel onto the canvas becomes a page pane. It shows the source page as that page draws it, by reference: edits to the source appear in the pane. A page cannot contain itself or a page that contains it. If the source page is deleted, the pane stays and reads "Page not found".
- Artwork opens the annotation editor over the whole page: text, arrows, lines, rectangles, ellipses, polygons and structures. A structure comes from the components of the displayed sample or from the clipboard.
- Moving, resizing, adding, removing and reordering panes are undone with Mod+Z; artwork edits use the editor's own undo.
In reports, {{page.figure}} draws the whole canvas as one picture: every pane with its own overlays, page panes drawn the same way, and the page artwork on top. Structures in the artwork are drawn by the backend.
Importing a .mnova document places its drawings (arrows, polygons, rectangles, text) on a free page of its spectra. The spectra are tiled; the drawings keep their arrangement and proportions and are scaled to fill the page.
Page layouts#
Page layouts... in the Pages panel's menu lists the layouts saved in the profile.
- Save pages as page layout... stores the open pages under a name. Each scan becomes a slot that matches an experiment type and nucleus, and keeps its colour, line width and layers; the page keeps its arrangement, scaling, spacing, plot style and window. Saving under an existing name replaces it.
- Choosing a layout places the scans of the working set in its slots: a slot takes the first scan that matches, in order. A slot with no matching scan is reported and left empty, and scans no slot takes get pages of their own.
- The star marks the layout that new samples get. When scans are added to a project that has no pages, they are placed by that layout; scans that arrive later fill the slots still open until the pages are edited.
- In a report,
{{page.figure|layout:Name}}draws the page's scans as the named layout arranges them, so a template with a repeated page block draws the same figure for any sample.
Display settings#
The plot's own settings panel controls presentation rather than analysis.
- General — plot title, and the title and type (linear, log, date, category) of each axis.
- Legend — visibility and orientation.
- Display — theme, whether the plot is editable in place, axis drag handles, axis range entry boxes, and when the mode bar appears.
- Image export — format, scale, width and height for saved images.
- Interaction — scroll zoom, and what a double click does.
- Plot style — line width, tick label size, axis title size, font and grid, saved on the page that holds the scan; Project applies the same settings to every page that sets none of its own. A scan's own line width is set from its colour menu on the page legend. Reports draw pages with the same style.
- Data — per-trace name, type, colour, line width and marker size, plus the raw data table with a Download button.
Traces are named Re, Im and Abs for the real, imaginary and absolute parts of a complex signal.