NMR Glossary#
A practical reference for the NMR terms you'll encounter in TQT Nuclei. Each entry explains what the term means, where it appears in the app, and when it matters. The same entries open in the app's Glossary panel.
- Acquisition time
-
How long the receiver records the FID after the pulse. A longer acquisition captures more of the decay and gives finer digital resolution; stopping too early causes truncation artefacts.
\[ \mathrm{AQ} = \mathrm{TD} \cdot \Delta t = \frac{\mathrm{TD}}{\mathrm{SW}} \] - Aliasing (folding)
-
What happens to a line whose offset lies outside the spectral width: it is not lost but shows up at a wrong position inside the window. With quadrature detection it wraps in from the opposite edge; with a single receiver channel it is mirrored about the edge. Sampling at least twice per period of the highest frequency (the Nyquist condition) avoids it.
\[ \nu_\text{app} = \nu - k\,\mathrm{SW}, \quad k \in \mathbb{Z} \] - Apodization
-
A window function that multiplies the FID before the Fourier transform. It removes ringing and noise at the cost of slightly broader lines; the exponential window's strength is the line broadening.
\[ s'(t) = s(t) \cdot w(t) \] - Baseline correction
-
Removing the slow offset or curvature underneath a spectrum, so that regions without signal sit at zero. Integrals depend on it: an uncorrected baseline adds area under every line or takes it away.
\[ S_\text{corrected}(\nu) = S(\nu) - b(\nu) \] - Chemical shift (δ, ppm)
-
Where a line sits relative to a reference compound, in parts per million of the spectrometer frequency. It does not depend on the field strength, so shifts compare across instruments; referencing sets its zero.
\[ \delta = \frac{\nu - \nu_\text{ref}}{\nu_0} \times 10^{6} \] - CPMG (Carr-Purcell-Meiboom-Gill)
-
A pulse sequence that measures T2 from a train of spin echoes, each refocused by a 180° pulse. It cancels field inhomogeneity better than a single Hahn echo; run it from the Acquisition page or the Spin-Spin Relax. page.
\[ A_n = A_0 \, e^{-2 n \tau / T_2} \]Tip: More echoes give a more accurate T2 curve, but very long echo trains can accumulate errors. Start with the default and adjust from there.
- Dead time (ring-down time)
- The delay after a radiofrequency pulse before the receiver starts recording, while the probe is still ringing from the pulse. Too short records hardware noise; too long loses the start of the FID and needs more phase correction. Tune it in the Dead Time tab of the Home page.
- Decimation
- Skipping a set number of 180° pulses between recorded echoes in a CPMG train. It samples late echoes without recording every one, which suits samples with a long T2. Set it as Decimation on the Spin-Spin Relax. page, or in the CPMG sequence parameters.
- Digital resolution
-
The frequency spacing between neighbouring points of a spectrum, in Hz per point: the spectral width divided by the number of points. Zero filling makes the spacing finer, but lines closer together than about 1 / acquisition time stay unresolved.
\[ \Delta \nu = \frac{\mathrm{SW}}{\mathrm{TD}} \] - Dwell time
-
The interval between consecutive data points during acquisition (typically 0.5 µs). It is set by the hardware and fixes the spectral width; it is shown in Settings → Device → Dwell Time.
\[ \Delta t = \frac{1}{\mathrm{SW}} \] - Echo time (τ)
- The delay between the excitation pulse (p90) and the refocusing pulse (p180) in an echo sequence such as the Hahn echo or CPMG. The echo forms at 2τ, so τ decides which part of the T2 decay you sample.
- Ernst angle
-
The flip angle that gives the most signal per scan when scans are repeated faster than full T1 recovery. It is shorter than 90° whenever the repetition delay is short; it maximises sensitivity, but integrals of lines with different T1 are then not quantitative.
\[ \cos\alpha_\text{E} = e^{-T_R/T_1} \] - FID (Free Induction Decay)
-
The time signal the excited nuclei emit after a single radiofrequency pulse: an oscillation at each line's frequency, dying away as the nuclei lose coherence. The Fourier transform turns it into a spectrum.
\[ s(t) = \sum_k A_k \, e^{i 2 \pi \nu_k t} \, e^{-t / T_{2,k}^{*}} \] - FIR filter (Finite Impulse Response)
- A digital filter applied to the raw signal to reduce noise. Choose between Gauss (smoother) and Sync (sharper cutoff) in Settings → Device → FIR Type, or turn on Bypass FIR if you need raw data.
- Fourier transform (FT, FFT)
-
The calculation that turns the time-domain FID into a frequency-domain spectrum: each frequency present in the signal becomes a line. The fast Fourier transform (FFT) is the algorithm that computes it.
\[ S(\nu) = \int_{0}^{\infty} s(t) \, e^{-i 2 \pi \nu t} \, \mathrm{d}t \] - Frequency / Larmor frequency
-
The frequency at which nuclei precess in the magnetic field; the device must be tuned to it to detect signal. Find it in the Frequency tab of the Home page, or set it in Settings → Device → Frequency.
\[ \nu_0 = \frac{\gamma}{2 \pi} B_0 \]Tip: If you're getting no signal, the frequency is the first thing to check. Use the auto-tune feature to find it automatically.
- Hahn echo
- A two-pulse sequence (90°–τ–180°) that produces a single spin echo at 2τ. It is simpler than CPMG and measures T2 when you don't need a full echo train; run it from the Acquisition page.
- Impurity table (known impurities)
- The ¹H and ¹³C positions of common laboratory solvents, reagents and greases in each deuterated solvent, from Fulmer et al., Organometallics 2010, 29, 2176 (298 K, dilute). A reported signal that sits on every line of a tabulated compound, within a window set by the spectrum's own line width, is noted with that compound's name. The note is advisory: the signal keeps its tier, because an analyte line can sit there too. Solvents the paper does not cover (pyridine-d5, DMF-d7, acetic acid-d4, TFA-d, dioxane-d8) have only the grease lines.
- Atom numbering (locants)
- How atoms are named in assignments, tables, reports and the structure overlay (Settings › Appearance › Atom numbering). Conventional uses a structure file's or drawing's own numbers where there are any; otherwise the established locants of the parent ring system (caffeine: N1, C2, N3, C4, C5, C6, N7, C8, N9, with its methyls as N1-CH₃, N3-CH₃, N7-CH₃ and its carbonyl oxygens O2 and O6) and numbered chains (from the end nearest the most senior group; ethylene glycol: C1, C2, O1, O2). A structure no rule covers is labelled a1, a2, … and marked as such. Each component of a mixture is numbered on its own, and every component after the first is primed (C1′). The ring systems are purine, pyrimidine, pyridine, benzene, naphthalene, indole, quinoline and the pyranose and furanose rings; As in the structure file and InChI canonical numbering remain as options. Tables and text write the element with the locant (C2, N1-CH₃; protons H-8). A drawing writes the same locant bare: ring positions as 1–9, substituents by the position they sit on (1-CH₃), and an atom without hydrogens that carries no assignment (a carbonyl oxygen) unlabelled unless every atom is requested. The app-wide setting Settings › Appearance › Atom numbering switches between Conventional (ring locants), as described above, and Sequential (structure order). Sequential labels every atom with its 1-based position in the structure's own order, written as the bare number (5) in drawings, assignment tables, copied peak lists and journal text, reports and exports; the protons on an atom take the number of that atom. The structure's own order is the atom order of the structure file where the sample kept it, and otherwise the order of the stored structure. Each mixture component is numbered from 1, and components after the first are primed. The setting changes labels only; assignments are stored by atom index and are unchanged by it. The project's own numbering source (Settings › Appearance › Numbering source for conventional labels) applies to conventional labels only.
- Integral
-
The area under a line or multiplet of a spectrum. It is proportional to the number of nuclei the signal comes from, which is why integrals, not heights, are compared in quantitative NMR. It needs a flat baseline, correct phase and full relaxation between scans.
\[ I = \int_{\nu_1}^{\nu_2} S(\nu)\,d\nu \;\propto\; N_\text{nuclei} \] - Inversion recovery
-
A pulse sequence (180°–τ–90°) that measures T1: the 180° pulse inverts the magnetization, and the 90° pulse reads how much has recovered after a variable wait τ. Used by the Spin-Lattice Relax. page.
\[ M_z(\tau) = M_0 \left(1 - 2 e^{-\tau / T_1}\right) \] - J-coupling (scalar coupling)
-
The interaction between nuclear spins through the bonding electrons, which splits each line into a multiplet. Its size J is given in Hz and does not depend on the field strength, unlike the chemical shift difference in Hz.
\[ \hat{H}_J = J_{12}\,\hat{\mathbf{I}}_1\cdot\hat{\mathbf{I}}_2 \] - Journal style
- How a journal writes an NMR experimental line: where the nucleus and δ sit,
J = 7.1 HzorJ 7.1,2Hor2 H, the number of decimals. The Report Builder and the peak table's copy button use one table of styles, so a copied line and a report agree. - Line broadening (LB)
-
The width in Hz that an exponential apodization window adds to every line. Larger values give less noise and broader, less resolved lines; a value close to the natural line width gives the best signal-to-noise ratio.
\[ w(t) = e^{-\pi \, \mathrm{LB} \, t} \] - Linear prediction (LP)
- Extending a truncated FID, or replacing its first points, with values predicted from the points that were acquired, before apodization and the Fourier transform. Forward prediction lengthens an FID that was cut off before it decayed, which reduces the sinc wiggles around lines; backward prediction replaces the first few points. It adds no information beyond the fitted model, and it is off unless switched on in Settings → Preprocessing → Linear Prediction or in the plot's Processing panel.
- Line-shape correction (reference deconvolution)
- Removing the instrument's lineshape distortion from every line by dividing one clean reference line out of the FID and multiplying an ideal line of a chosen width in. It acts on the FID, before apodization, so it is not the same as deconvolving the spectrum afterwards. It is off unless switched on in Settings → Preprocessing → Line-shape correction. Because one line decides the shape of all the others, qNMR warns when it has been applied.
- Linewidth (lineshape)
-
The full width of a line at half its height (FWHM), in Hz, and the shape of the line around it. A single exponential decay gives a Lorentzian whose width is set by T2*; poor shimming broadens and distorts it, and line broadening adds to it.
\[ \Delta\nu_{1/2} = \frac{1}{\pi T_2^*} \] - Multiplet (n + 1 rule)
-
The group of lines a signal splits into through J-coupling. Coupling to n equivalent spin-½ neighbours gives n + 1 lines, J apart, with intensities from Pascal's triangle (1 : 2 : 1, 1 : 3 : 3 : 1 …). The rule holds only when the shift difference is much larger than J; otherwise second-order effects appear.
\[ I_k = \binom{n}{k}, \quad k = 0 \ldots n \] - Net magnetization
-
The vector sum M of all nuclear magnetic moments in the sample. At equilibrium it points along the field with size M0; pulses tip it (nutation), it then precesses at the Larmor frequency, and T1 and T2 return it to equilibrium. Its transverse part is what the receiver detects.
\[ \frac{d\mathbf{M}}{dt} = \gamma\,\mathbf{M}\times\mathbf{B} \] - Nutation
-
Sweeping the pulse duration and watching the signal to find the optimal flip angle (p90): the signal peaks at 90° and crosses zero at 180°. Performed in the Duration tab of the Home page.
\[ S(t_p) \propto \sin\left(\gamma B_1 t_p\right) \] - Phase correction
-
Adjusting the zero-order phase (p0, the same at every frequency) and the first-order phase (p1, changing linearly across the spectrum) until every line is upright and symmetric. A line left out of phase has a dispersive, partly negative shape and integrates wrongly.
\[ \varphi(\nu) = \varphi_0 + \varphi_1 \cdot \nu \] - Phase cycling
- Varying the phase of the radiofrequency pulses across scans so that artefacts cancel in the sum. Enable it with Phase Cycling on the Acquisition page; the number of scans must then be a multiple of 4.
- Pulse duration (p90, p180, p270, p360)
-
The length of the radiofrequency pulse in microseconds, which decides how far the magnetization is tipped: p90 tips it 90° (maximum signal), p180 inverts it. The values depend on the sample; find them with nutation in the Duration tab of the Home page.
\[ \theta = \gamma B_1 t_p \]Tip: Re-tune pulse durations when changing samples or after significant temperature changes.
- Quadrature detection
-
Recording the signal on two receiver channels 90° apart, giving a complex FID. It tells positive offsets from negative ones, so the carrier can sit in the middle of the spectral width. Mismatched channels leave a mirror image that phase cycling removes.
\[ s(t) = s_x(t) + i\,s_y(t) \] - Quantitative NMR (qNMR)
-
Measuring amounts or purity from integrals, against a reference of known amount. It needs every signal fully relaxed before each scan (a repetition delay of about 5–7 × the longest T1), enough signal-to-noise, and careful phasing and baseline correction.
\[ \frac{n_a}{n_\text{ref}} = \frac{I_a / N_a}{I_\text{ref} / N_\text{ref}} \] - Referencing
- Setting the ppm axis so that a known signal sits at its accepted chemical shift, for example TMS at 0 ppm or residual CHCl₃ at 7.26 ppm. Referencing moves the axis only; line shapes and areas do not change.
- Repetition delay
-
The wait between consecutive scans, which lets the magnetization recover. It should be at least 5 × T1, which recovers more than 99 %; the software sets it from the expected T1, and you can override it in Settings → Device → Repetition Delay.
\[ \frac{M_z}{M_0} = 1 - e^{-d_1 / T_1} \] - Ringing / truncation artefacts
- Ripples on both sides of a line (sinc wiggles), caused by cutting the FID off before it has decayed. A longer acquisition time or an apodization window that tapers the end of the FID removes them.
- Rotating frame
-
A coordinate frame that turns about the field at the carrier frequency. Seen from it, the fast Larmor precession disappears: a line on resonance stands still, a line off resonance turns slowly at its offset, and a pulse is a static field about which the net magnetization nutates.
\[ \Omega = \omega_0 - \omega_\text{rf} \] - Second-order effects (roofing)
-
Distortions of multiplets when the shift difference Δν between coupled groups is not much larger than J. Lines lean towards the coupling partner (the roof effect), their spacings shift and extra lines appear. They fade as Δν/J grows, so a higher field gives simpler spectra.
\[ \frac{I_\text{inner}}{I_\text{outer}} = \frac{\sqrt{\Delta\nu^2 + J^2} + J}{\sqrt{\Delta\nu^2 + J^2} - J} \] - Shielding (σ)
-
The reduction of the field a nucleus feels caused by its surrounding electrons. More shielded nuclei resonate at lower frequency and lower chemical shift; electronegative neighbours remove electron density and deshield.
\[ \nu = \frac{\gamma B_0}{2\pi}(1 - \sigma) \] - Shimming
-
Adjusting small correction coils (shims) until the field is the same throughout the sample. A linear error (Z1) broadens a line symmetrically; a quadratic one (Z2) makes it lopsided. Good shims give narrow, symmetric lines and a short T2* contribution; see linewidth.
\[ \Delta B_0(z) = Z_1\,z + Z_2\left(z^2 - \tfrac{1}{3}\right) + \ldots \] - Signal averaging (number of scans)
-
Adding N repeated scans. The signal grows as N and random noise as √N, so the signal-to-noise ratio grows as √N: twice the signal-to-noise costs four times the measurement time. Scans should come in whole phase cycles.
\[ \mathrm{SNR}_N = \sqrt{N}\,\mathrm{SNR}_1 \] - Signal vs. Spectrum
- Signal is the raw time-domain data (FID, amplitude over time); Spectrum is the frequency-domain data after the Fourier transform (amplitude over frequency). Switch between them with the Signal and Spectrum tabs of the plot on the Acquisition page.
- Signal-to-noise ratio (SNR)
-
How tall a line is compared with the noise in a region without signal. It grows with the square root of the number of scans, so four times as many scans doubles it.
\[ \mathrm{S/N} = \frac{h_\text{line}}{\sigma_\text{noise}} \propto \sqrt{N_\text{scans}} \] - Spectral width (SW)
-
The range of frequencies a spectrum covers, set by the dwell time. Lines outside it fold back into the spectrum at the wrong position.
\[ \mathrm{SW} = \frac{1}{\Delta t} \] - T1 (spin-lattice relaxation time)
-
The time constant with which the magnetization recovers along the field after a pulse; it decides how long to wait between scans (see repetition delay). Measured on the Spin-Lattice Relax. page with inversion recovery.
\[ M_z(t) = M_0 \left(1 - e^{-t / T_1}\right) \] - T2 (spin-spin relaxation time)
-
The time constant with which the transverse magnetization decays after excitation; a shorter T2 means a faster-fading signal and broader lines. Measured on the Spin-Spin Relax. page with CPMG.
\[ M_{xy}(t) = M_0 \, e^{-t / T_2} \]Practical meaning: T2 tells you about molecular mobility. Liquids have a long T2 (slow decay), solids a short one (fast decay). Water at room temperature: T2 ≈ 2–3 seconds.
- T2 (effective transverse relaxation)*
-
The time constant with which the FID actually decays: true T2 loss plus dephasing from field inhomogeneity. The inhomogeneous part can be refocused by a Hahn echo or CPMG; it sets the observed linewidth.
\[ \frac{1}{T_2^*} = \frac{1}{T_2} + \pi\,\Delta\nu_\text{inh} \] - Zero filling
- Adding zeros to the end of the FID before the Fourier transform. It adds no information but interpolates between frequency points, so the digital resolution is finer and lines look smoother. Settings → Preprocessing → Zero Filling turns it on and sets the factor.